Loading files…
Firmware version 1.0.2 for HHS's portable industrial gas monitor. The application runs on a custom nRF52832 board, measures multiple gases, and transmits real-time readings over Bluetooth® Low Energy. Environmental data is fused with the Bosch BSEC2 library to provide indoor air-quality indicators in addition to electrochemical gas concentrations.
boards/arm/hhs_nrf52832 (schematic and layout
included in doc/).【F:boards/arm/hhs_nrf52832/hhs_nrf52832.yaml†L1-L53】doc/gas_nrf52832_rev1.0.1.pdf with matching
schematic doc/gas_nrf52832_sch_rev0.1.0.pdf.| Domain | Reported values |
|---|---|
| Electrochemical gas | Oxygen concentration (%VOL) and one selectable toxic gas in |
| ppm, both temperature compensated and filtered.【F:src/gas.c†L1-L212】 | |
| Environmental | Temperature (°C), relative humidity (%RH), barometric |
| pressure (Pa), IAQ score, equivalent CO₂ (ppm), and breath | |
| VOC index via BSEC2.【F:drivers/bme68x_iaq/bme68x_iaq.c†L40-L460】 | |
| Power | Battery percentage derived from loaded voltage profile. |
| BLE telemetry | Timestamped payload streamed every 10 seconds (default). |
The firmware is organized into cooperative Zephyr threads and kernel events. A high-level flow is shown below:
---
title: Portable Gas Monitor Firmware Flow
---
stateDiagram-v2
direction TB
[*] --> ZephyrRTOS
ZephyrRTOS --> BME68x_BSEC
ZephyrRTOS --> Electrochemical_ADC
ZephyrRTOS --> BluetoothLE
state ZephyrRTOS {
state BME68x_BSEC {
direction LR
initBSEC --> readEnvironmental : 3 s
initBSEC --> computeIAQ : 60 s
}
state Electrochemical_ADC {
direction TB
sampleADC --> tempCompensation
tempCompensation --> gasConversion
gasConversion --> dynamicCalibration
}
state BluetoothLE {
direction LR
initialize --> advertise
advertise --> connected
connected --> notifyData : 10 s
}
}
Key modules reside in src/ (application logic), drivers/bme68x_iaq/
(BSEC2 integration), and lib/ (Bosch libraries). Threads communicate via
Zephyr events for deterministic BLE publishing and alarm handling.
west, and SEGGER J-Link tools.lib/.git clone https://github.com/<your-org>/gas_nordic.git
cd gas_nordic
west init -l .
west update
west build -b hhs_nrf52832 .
The output image is generated at build/zephyr/zephyr.hex.west flash
or, to use Nordic's command-line tools directly:
./flash.sh
(flash.sh performs a chip erase and programs the generated HEX via
nrfjprog).【F:flash.sh†L1-L5】The device exposes a proprietary 128-bit primary service (0000FFF0-0000-1000-8000-00805F9B34FB).【F:src/bluetooth.h†L9-L26】
| Characteristic | UUID | Properties | Description |
|---|---|---|---|
| Measurement | 0000FFF1-0000-1000-8000-00805F9B34FB | Notify | Periodic payload O2;Gas;Battery;Temp;Pressure;Humidity separated by semicolons, integers scaled as <val1>.<val2> where applicable.【F:src/bluetooth.c†L536-L576】 |
| Command | 0000FFF2-0000-1000-8000-00805F9B34FB | Write | ASCII commands for calibration and configuration: O2=<percent>, NO2=<ppm>, BT=<name>.【F:src/bluetooth.c†L60-L124】 |
Notifications are issued when a connection is active and the client enables CCCD. Each update corresponds to the latest sensor snapshot and is throttled to the configured 10 s publishing interval.
src/battery.c).| Category | Concentration range | Notes |
|---|---|---|
| High oxygen | > 24 % | Elevated combustion risk. |
| Normal air | 20.9 % | Nominal breathable atmosphere. |
| Elevated exposure risk | 19.5 – 20.9 % | Monitor for dilution hazards. |
| Low oxygen | 16 – 19.5 % | Impaired cognition and physical performance. |
| Anoxic | < 16 % | Life-threatening within minutes. |
| IAQ index | Air quality | Suggested action |
|---|---|---|
| 0 – 50 | Excellent | None required. |
| 51 – 100 | Good | None required. |
| 101 – 150 | Lightly polluted | Ventilate if convenient. |
| 151 – 200 | Moderately polluted | Increase ventilation with clean air. |
| 201 – 250 | Heavily polluted | Identify contamination source, improve airflow. |
| 251 – 350 | Severely polluted | Restrict access, maximize ventilation. |
| > 351 | Extremely polluted | Avoid occupancy until levels normalize. |
| Level | Description | ppm |
|---|---|---|
| 😆 | Healthy outdoor reference | 350 |
| 😀 | Healthy indoor climate | 600 |
| 🙂 | Acceptable indoors | 800 |
| 🙁 | Ventilation desirable | 1000 |
| 😩 | Ventilation necessary | 1200 |
| 😡 | Noticeable health effects | 2000 |
| ⚠️ | Dangerous for prolonged exposure | 5000 |
| Rating | Recommendation | Exposure duration | TVOC (ppm) |
|---|---|---|---|
| Excellent | Target value | Unlimited | 0 – 0.065 |
| Good | Ventilation recommended | Unlimited | 0.065 – 0.22 |
| Moderate | Increase ventilation, inspect for sources | < 12 months | 0.22 – 0.66 |
| Poor | Intensify ventilation, locate emission sources | < 1 month | 0.66 – 2.2 |
| Unhealthy | Immediate ventilation required | Hours | 2.2 – 5.5 |
Distributed under the terms of the MIT License. Bosch BSEC2 library
usage is subject to its proprietary license (see lib/Bosch-BSEC2-Library).